Analysis of the Effect of Varying Slice Thickness on Image Quality in CT Scans of the Brain With Clinical Infarction at Siti Rahmah Islamic Hospital in Padang

Authors

  • Wahdini Hanifah DIII Radiology Study Program, Vocational Faculty, Baiturrahmah University, Indonesia
  • Nerifa Dewilza DIII Radiology Study Program, Vocational Faculty, Baiturrahmah University, Indonesia
  • Sagita Yudha DIII Radiology Study Program, Vocational Faculty, Baiturrahmah University, Indonesia
  • Livia Ade Nansih DIII Radiology Study Program, Vocational Faculty, Baiturrahmah University, Indonesia
  • Santa Mareta DIII Radiology Study Program, Vocational Faculty, Baiturrahmah University, Indonesia
  • Rina Astuti DIII Radiology Study Program, Vocational Faculty, Baiturrahmah University, Indonesia

DOI:

https://doi.org/10.65739/archipel.v1i10.65

Keywords:

CT-Scan, Slice Thickness, Image Quality, Cerebral Infarction, CNR, SNR

Abstract

An Analysis Of The Effect Of Slice Thickness Variation On Image Quality In Ct-Scan Examination Of The Brain With Clinical Infarction At Siti Rahmah Islamic Hospital In Padang. Infarction stroke occurs when blood flow in the brain is disrupted, causing damage to brain tissue. The appropriate slice thickness increases picture quality and allows for more accurate detection of infarction stroke, reducing the possibility of error. The purpose of this study is to investigate the influence of slice thickness variation on image quality in a brain CT-scan examination with clinical infarction performed at RSI Siti Rahmah Padang. This study was carried out at the RSI Siti Rahmah Padang radiology installation in May-June 2024 utilizing an experimental quantitative research design and a purposive sample technique. Five CT-Scan infarction pictures were taken with a slice thickness of 5 mm and reconstructed to 2 mm and 3 mm. The reconstruction result is then in the ROI on the radiant dicom, then the CNR and SNR calculations are carried out and analysed using SPSS with the Manova test. The result of the slice thickness variation value towards CNR and SNR that has been processed using the Manova test obtained R square CNR 0.839 and SNR 0.797 with a significance level of p of 0.000<0.05, then it was concluded that the slice thickness variation had an effect on the CNR and SNR values. The highest mean value that is optimally meaningful is obtained at a slice thickness of 5 mm, so it can be concluded that the size of the slice thickness is directly proportional to the increase in CNR and SNR values.       

Downloads

Download data is not yet available.

References

Abdulkareem, N., Hajee, S., Hassan, F., Ibrahim, I. K., Al-Khalidi, R. E. H., & Abdulqader, N. A. (2023). Investigating the slice thickness effect on noise and diagnostic content of single-source multi-slice computerized axial tomography. Journal of Medicine and Life, 16, 862–867. https://doi.org/10.25122/jml-2022-0188

Anggarani, N. K. N., Suryatika, I. B. M., & Sutapa, G. (2023). Effect of slice thickness variation on spatial resolution of CT-scan device images at the Bali Mandara Hospital Radiology Installation. Asian Journal of Medicine and Health. https://doi.org/10.9734/ajmah/2023/v21i10901

Baert, A. (2018). Multislice CT (4th ed.). Springer Heidelberg Dordrecht.

Choopani, M., Abedi, I., & Dalvand, F. (2022). Pragmatic approaches to reducing radiation dose in brain computed tomography scan using scan parameter modification. Journal of Medical Signals and Sensors, 12, 219–226. https://doi.org/10.4103/jmss.jmss_83_20

Dewilza, N., Yudha, S., Hanifah, W., Nansih, L. A., & Septiani, R. (2024). Analysis of differences in image quality with variations in slice thickness in CT-scan brain examinations with trauma cases at the Radiology Installation of RSI Siti Rahmah Padang. Jurnal Ilmu dan Teknologi Kesehatan. https://doi.org/10.33666/jitk.v15i2.643

Funama, Y., Shirasaka, T., Goto, T., Aoki, Y., Tanaka, K., & Yoshida, R. (2022). Iterative reconstruction with multifrequency signal recognition technology to improve low-contrast detectability: A phantom study. Acta Radiologica Open, 11. https://doi.org/10.1177/20584601221109919

Harkey, T., Baker, D., Hagen, J., Scott, H., & Palys, V. (2021). Practical methods for segmentation and calculation of brain volume and intracranial volume: A guide and comparison. Quantitative Imaging in Medicine and Surgery, 12(7), 3748–3761. https://doi.org/10.21037/qims-21-958

Hirairi, T., Ichikawa, K., Urikura, A., Kawashima, H., Tabata, T., & Matsunami, T. (2023). Improvement of diagnostic performance of hyperacute ischemic stroke in head CT using an image-based noise reduction technique with non-black-boxed process. Physica Medica, 112, 102646. https://doi.org/10.1016/j.ejmp.2023.102646

Hutami, I. A. P. A., Sutapa, G., & Paramarta, I. (2021). Analisis pengaruh slice thickness terhadap kualitas citra pesawat CT scan di RSUD Bali Mandara. Buletin Fisika, 22(2), 77–83.

Inoue, Y., Itoh, H., Hata, H., Miyatake, H., Mitsui, K., Uehara, S., & Masuda, C. (2024). Noise reduction in brain CT: A comparative study of deep learning and hybrid iterative reconstruction using multiple parameters. Tomography, 10, 2073–2086. https://doi.org/10.3390/tomography10120147

Karina. (2017). Analisis pengaruh variasi slice thickness terhadap kualitas citra CT scan. [Thesis, Indonesia].

Maghfirah, A., Rahim, A., & Fadhlin, A. (2024). Linearity test of CT number and spatial resolution using ACR 464 phantom for image quality assessment on a 16-slice CT scanner. Journal of Technomaterial Physics. https://doi.org/10.32734/jotp.v6i2.12410

Mitani, H., Tatsugami, F., Higaki, T., Kaichi, Y., Nakamura, Y., Smit, E., Prokop, M., Ono, C., Ono, K., Korogi, Y., & Awai, K. (2021). Accuracy of thin-slice model-based iterative reconstruction designed for brain CT to diagnose acute ischemic stroke in the middle cerebral artery territory: A multicenter study. Neuroradiology, 63, 2013–2021. https://doi.org/10.1007/s00234-021-02745-4

Nelson, B., Gomez-Cardona, D., Thorne, J., Huber, N., Yu, L., Leng, S., McCollough, C., & Missert, A. (2024). Multiple kernel synthesis of head CT using a task-based loss function. Journal of Imaging Informatics in Medicine, 37, 864–872. https://doi.org/10.1007/s10278-023-00959-x

Notoatmojo, S. (2018). Metodologi penelitian kesehatan. Rineka Cipta.

Peng, Y., Luo, C., Wang, H., Sun, K., Lin, F., Wang, J., Rao, Y., Fan, R., Gong, L., & Sun, X. (2024). Feasibility of CT attenuation values in distinguishing acute ischemic stroke, old cerebral infarction and leukoaraiosis. BMC Medical Imaging. https://doi.org/10.1186/s12880-024-01340-2

Putu, L., Kusumaningsih, R., Suryatika, I. B. M., Trisnawati, N. L. P., & Irhas, R. (2023). Pengaruh slice thickness terhadap signal to noise ratio (SNR) dari hasil penyinaran CT Scan di RSUP Prof. Dr. I.G.N.G Ngoerah. Kappa Journal. https://doi.org/10.29408/kpj.v7i2.7675

Rajendran, K., Petersilka, M., Henning, A., Shanblatt, E., Marsh, J., Thorne, J., Schmidt, B., Flohr, T., Fletcher, J., McCollough, C., & Leng, S. (2021). Full field-of-view, high-resolution, photon-counting detector CT: Technical assessment and initial patient experience. Physics in Medicine and Biology, 66. https://doi.org/10.1088/1361-6560/ac155e

Rinaldi, S. F., & Mujianto, B. (2017). Metodologi penelitian dan statistik. Trans Info Media.

Sato, T., & Takagi, T. (2021). Validation of the influence of CT slice thickness on the quantitative accuracy and image quality of single photon emission computed tomography. Asia Oceania Journal of Nuclear Medicine and Biology, 9(2), 148–157. https://doi.org/10.22038/aojnmb.2021.54670.1376

Seeram, E. (2016). Computed tomography: Physical principles, clinical applications, and quality control (4th ed.). Elsevier.

Siregar, E. S. (2020). Analisis dosis radiasi pasien pada pemeriksaan CT scan menggunakan aplikasi Si-INTAN. Buletin Fisika, 21(2), 53–59.

Solong, D. I., Putu, N., Jeniyanthi, R., & Sukadana, I. (2023). Analisis pengaruh variasi slice thickness terhadap kualitas citra pada pemeriksaan CT scan kepala dengan kasus stroke hemorragic di RS TK II Pelamonia Makassar. Journal of Educational Innovation and Public Health. https://doi.org/10.55606/innovation.v1i4.1992

Sun, J., Li, H., Wang, B., Li, J., Li, M., Zhou, Z., & Peng, Y. (2021). Application of a deep learning image reconstruction (DLIR) algorithm in head CT imaging for children to improve image quality and lesion detection. BMC Medical Imaging, 21. https://doi.org/10.1186/s12880-021-00637-w

Sun, J., Yao, H., Han, T., Wang, Y., Yang, L., Hao, X., & Wu, S. (2025). Thin-slice brain CT image quality and lesion detection evaluation in deep learning reconstruction algorithm. Clinical Neuroradiology, 35, 817–825. https://doi.org/10.1007/s00062-025-01542-3

Vromans, J., Bennink, E., Dankbaar, J., Velthuis, B., & De Jong, H. W. A. M. (2026). Robust filtering of thin-slice reconstructions improves lacunar stroke detection in CT perfusion imaging. Medical Physics, 53. https://doi.org/10.1002/mp.70271.

Downloads

Published

29-06-2024

How to Cite

Hanifah, W., Dewilza, N., Yudha , S., Nansih, L. A., Mareta, S., & Astuti, R. (2024). Analysis of the Effect of Varying Slice Thickness on Image Quality in CT Scans of the Brain With Clinical Infarction at Siti Rahmah Islamic Hospital in Padang. Archipel: Journal of Indonesian Interdisciplinary Studies, 1(10), 25–34. https://doi.org/10.65739/archipel.v1i10.65